If the EH (electrode heating) supply is damaged or we need to replace it for some other reasons, please follow this information carefully!

Location of the EH supply
Location of the EH supply


  1. Instructions
  • Take out the 230V fuse and wait 5 min if the generator was switched on before to let the capacitors discharge
  • Switch off the 24V supply of the generator
  • Remove the polarity switch knobs.
  • Remove the plexi glass cover, one screw is behind the warranty seal of Heinzinger. It is okay to break the seal in this case, but we need to inform Heinzinger later on
  • Write the date and your name on the HV cascade as well as the reason you take it out
  • unscrew the current sensor at the Supply
Location of the EH current sensor
Location of the EH current sensor
  • Mark the HV cable connections with a marker or a piece of tape. The connections are all pluged in, but we will get problems if they are not fully inserted. To avoid this, this marking will help you later to check if they are fully plugged in again.
Marking of EH cables at the connections left side
Marking of EH cables at the connections left side
Marking of EH cables at the connections right side
Marking of EH cables at the connections right side
EH connector top
EH connector top
  • Unscrew the supply, take it out carefully and mount the new one
  • Plug in fully all connectors of the supply and mount the current sensor again
  • Reinstall the cover
  • When reinstalling the polarity switches, make sure they point to the correct direction. You can see the polarity at the small limit switches activated by the screw at the shaft.
File:E27 Polarity switches.png
Polarity switches
File:E27 Find correct position of polarity switches.png
Find correct position of polarity switches
  • perform a function test